Abstract

Effects of analog high pass (HP) filtering were compared with those of zero phase-shift digital filtering on the auditory middle latency responses (MLR) from nine adults and 16 young children with normal hearing. Analog HP filtering exerted several prominent effects on the MLR waveforms in both adults and young children, such as suppression of Po (ABR), enhancement of Nb, enhancement or emergence of Pb, and latency decrements for Pa and the later components. Analog HP filtering at 20 Hz produced more pronounced waveform distortions in the responses from young children than from adults. Much greater latency decrements for Pa and Nb were observed for young children than for adults in the analog HP-filtered responses at 20 Hz. A large positive peak (Pb) emerged at about 65 ms after the stimulus onset. From these results, the use of digital HP filtering at 20 Hz is strongly recommended for obtaining unbiased and stable MLR in young children.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.