Abstract

This study aimed to investigate the effects of alcohol-based hand hygiene solution (ABHS) use by care providers on point-of-care alcohol breath analyzer interpretation under different clinically relevant conditions. Among each test condition (foam vehicle with immediate testing, gel vehicle with immediate testing, allowing hands to dry after the use of ABHS, and donning gloves after the use of ABHS), alcohol was detected in breath at 1 minute after use of ABHS. Because the use of ABHS by individuals administering breath alcohol detection may result in false-positive detection of alcohol, staff using these devices should consider traditional hand hygiene with soap and water.