Abstract

To compare the effect of different acquisition and reconstruction methods on the radiation dose and accuracy of CT number measurements, using a 320-detector row CT and a Quantitative Imaging Biomarker Alliance (QIBA) recommended phantom. Acquisitions were performed on a 320-detector row CT, as 64- and 80-detector row helical and wide detector step-and-shoot (i.e., wide volume) acquisitions with tube currents of 400mA, 100mA, 50mA, 20mA, and 10mA. Image was reconstructed with the filtered back projection (FBP), adaptive iterative dose reduction using 3D processing (AIDR 3D), and forward projected model-based iterative reconstruction (FIRST) methods. The difference between measured CT numbers and the actual -856HU value of the phantom insert was determined by each CT acquisition protocol. Differences in actual and measured CT numbers were compared among acquisitions and among reconstruction methods by means of Tukey's HSD test. The CT number obtained with 64-detector row helical acquisition was significantly larger than that obtained with others (p < 0.0001). At each tube current, the CT number reconstructed with FIRST was significantly smaller than that with others (p < 0.0001). Acquisition and reconstruction methods are significantly affecting radiation dose reduction and accuracy of CT number measurements on a phantom study.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.