Abstract
Protective patient equipment for CT examinations is not routinely provided. The aim of this study was to determine whether, and if so what, specific protective equipment is beneficial during CT scans. The absorbed organ doses and the effective doses for thorax, abdomen/pelvis and brain CT investigation with and without the use of protective patient equipment have been determined and compared. All measurements were carried out on modern multislice CT scanner using an anthropomorphic phantom and thermoluminescence dosemeters. The measurements show that protective equipment reduces the dose within the scattered beam area. The highest organ dose reduction was found in organs that protrude from the trunk like the testes or the female breasts that can largely be covered by the protective equipment. The most reduction of the effective dose was found in the male abdomen/pelvis examination (0.32 mSv), followed by the brain (0.11 mSv) and the thorax (0.06 mSv). It is concluded that the use of protective equipment can reduce the applied dose to the patient.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.