Abstract

A novel approach using wavelength dispersive X-ray fluorescence (WDXRF) spectrometry combined with an activated carbon (AC) disk was developed for the determination of total iodine concentrations in water samples. Dissolved iodine species (i.e., I− and IO3−) in water samples were preconcentrated on the AC disk and directly analyzed by WDXRF spectrometry. The adsorption behavior of I− and IO3− on the AC disk was assessed at varying pH levels (4, 6, and 8). The AC disks completely retained the I− and IO3− for all the pH levels tested. The calibration curve obtained from the iodine concentrations (i.e., 0, 20, 200, and 400μg) of AC disks and the measured X-ray intensity from the WDXRF analysis showed a good linearity (R2=0.9960), with a relatively low limit of detection (0.575μg). The durability of the AC disk for repeatable measurements was also assessed to validate the sustainability of the proposed method and consequently the measured X-ray intensity for the AC disks was constant until 8d of analysis time. The accuracy of the proposed AC-WDXRF method was confirmed by measuring iodine concentration spiked in drinking water using inductively coupled plasma-mass spectrometry (ICP-MS). The proposed method is simple, rapid, efficient, and environmental friendly for iodine analysis in water samples. As a precursor of disinfection by products (DBPs), it is important to determine the total iodine concentrations in raw water.