Abstract

The clinical benefit of endovascular stroke therapy has been demonstrated in several prospective randomized trials. However, in a relevant percentage of patients, mechanical thrombectomy bears the risk of causing new infarction in initially unaffected vascular territories through thrombus fragmentation and migration of clot debris. The goal of this study was to evaluate the use of the balloon guide catheter (BGC) to effectively achieve flow arrest and thrombus aspiration during the intervention to avoid distal embolization. A retrospective study was performed in 139 patients between October 2010 and May 2016 to analyze occlusions in the middle cerebral artery (MCA) or internal carotid artery (ICA) by using a stent retriever with a BGC (n=73) or a non-BGC (n=66). The following data were collected: patient age and gender, along with history of diabetes mellitus, hypertension, atrial fibrillation, smoking, obesity, dyslipidemia, and previous ischemic stroke. Data on procedure time, number of passes, and angiographic findings were also collected. The final reperfusion score was rated based on the Thrombolysis in Cerebral Infarction (TICI) grading scale. Successful recanalization was defined as TICI 3 or 2b. A total of 139 patients underwent mechanical thrombectomy with the stent retriever. Of the 139 patients, 73 (52.5%) underwent placement of a BGC. The mean age was 65.8±13.5years, and the median National Institutes of Health Stroke Scale (NIHSS) score was 11. The average initial NIHSS score was lower in the BGC group compared with the non-BGC group (mean, 11.2±5.6 vs. 13.2±5.6; P=0.03). Patients with BGC had fewer incidences of previous ischemic stroke (12.3% vs. 28.8%; P=0.01). The numbers of passes were similar between the two groups. The procedure time (99±49.4min vs. 124±72.2min; P=0.02) and the time from onset of symptoms to procedure end (302±102min vs. 357.2±136.1min; P=0.009) were shorter in the BGC group. TICI 3 or 2b recanalization scores were higher in the BGC group compared to the non-BGC group [63/73, 86.3% vs. 48/66, 72.7%; odds ratio (OR), 0.6; 95% confidence interval (CI), 0.2-1.4; P=0.04]. Importantly, distal embolization was less frequent in the BGC group (5/73, 6.8% vs. 21/66, 31.8%; OR, 6.3; 95% CI, 2.2-18.0; P < 0.001). The risk of distal embolization was significantly decreased with the use of a BGC.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call