Abstract

Endobronchial radiofrequency ablation (RFA) is a novel minimally invasive approach to management of peripheral non-small-cell lung cancer (NSCLC) in medically inoperable patients. Minimally invasive ablative techniques are generally delivered with cone-beam computed tomography (CBCT) guidance. CBCT requires a significant number of two dimensional imaging projections to be acquired which is then reconstructed as a three-dimensional cone-beam image. The objective of this study was to determine the radiation dosimetry consequent to use of CBCT guidance for bronchoscopic RFA. Post hoc analysis of data following bronchoscopic RFA of stage I biopsy-confirmed NSCLC performed with CBCT. Effective dose estimates for these patients were calculated using PCXMC2.0 software. Ten patients underwent bronchoscopic RFA, with a median 3 (range 2-4) CBCT spins per procedure. Mean dose area product (DAP) per procedure was 7,778 μGy.m2 (±4,743) with an effective dose of 11.6 mSv (±7.4). The DAP per spin for these 10 patients varied from 83.8 to 8,625.6 μGy.m2 (effective dose range 0.15-13.81 mSv). This is the first study to report radiation dosimetry consequent to CT guidance for bronchoscopic RFA procedures. Effective doses appear comparable to other CT fluoroscopic procedures.