Abstract

We report a case of effective trans catheter repositioning of an ICD lead that was displaced during a trans venous extraction procedure of another malfunctioning ICD lead. This original technique was effective also in screwing-in the active fixation tip of the lead. Skilled operators could take into account this technique to avoid the re-opening of the device pocket, when dealing with specific situations at high risk of infection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.