Abstract
Purpose:The purpose of this study is to build a system for effective dose display immediately after the gastric cancer X-ray screening.Materials and Methods:The regression equation of effective dose and dose area product (DAP) was introduced from the data of 500 persons including DAP and effective dose calculated using program for X-ray Monte Carlo.Results:The effective dose was 5.39 mSv of median, 1.18 mSv of minimum, and 38.38 mSv of maximum. The regression equation was Y=0.354+0.0003772X (Y: effective dose, mSv, X: DAP, mGy cm2). Using the regression equation, the effective dose can be estimated from DAP and displayed just after the individual screening.Conclusions:“Effective dose display system” was constructed to display effective dose immediately after gastric cancer X-ray screening. This system is on the way to be reformed by improving the regression equation on larger data.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.