Abstract

Measurements of ultrasonic velocity and attenuation require ’’diffraction corrections’’ for the finite size of the source and receiver. These diffraction corrections depend on the area of the source which is usually taken as the physical area of the piezoelectric element of the the transducer. Measurements are reported which indicate that the effective area of a disk-shaped element may vary with the exciting frequency and may differ by 40% from the physical area of the disk. The method reported may be of value in the selection of transducers, in accurate diffraction correction calculations, and in the field of ultrasonic exposimetry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.