Abstract
A theoretical and experimental investigation of the influence of eccentricity of the multiplication wire on the performance of cylindrical proportional counters is presented. The electric field in the counter is calculated by the method of images, and the Townsend formalism is used to derive the gas gain. The experimental determination of detector performance is carried out with 37Ar. The dependence of the gas gain and of the counter resolution on eccentricity is discussed, and it is shown that eccentricities up to 0.2 are of no concern in microdosimetric measurements with cylindrical proportional counters.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.