Abstract

AbstractStandards for overhead emergency safety showers used for decontamination in laboratories and production sites specify the spray pattern and the water flow rate. Up to now, there has not been a scientific basis for these requirements. The decontamination of water‐soluble contaminants has been investigated using 15 customary shower heads at different water flow rates. A general relationship between water flow rate and decontamination effectiveness could not be established: At a low flow rate, the wash‐off time in general did not exceed the time observed at considerably higher flow rates. The shortest decontamination time has been found with a spray pattern that does not fulfil current standards.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.