Abstract
The passive fit between prosthesis and implant is a relevant factor for screw joint stability and treatment success. The aim of this study was to evaluate the influence of vertical misfit in abutment-implant interface on preload maintenance of retention screw of implant-supported crowns. The crowns were fabricated with different abutments and veneering materials and divided into 5 groups (n = 12): Gold UCLA abutments cast in gold alloy veneered with ceramic (Group I) and resin (Group II), UCLA abutments cast in titanium veneered with ceramic (Group III) and resin (Group IV), and zirconia abutments with ceramic veneering (Group V). The crowns were attached to implants by gold retention screws with 35-N cm insertion torque. Specimens were submitted to mechanical cycling up to 106 cycles. Measurements of detorque and vertical misfit in abutment-implant interface were performed before and after mechanical cycling. ANOVA revealed statistically significant difference (P 0.05) between vertical misfit and detorque value. It was concluded that vertical misfit did not influence torque maintenance and the abutments cast in titanium exhibited the highest misfit values.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.