Abstract

An Immersed Boundary (IB) model is employed to investigate the dynamic behaviour of a novel chorded mitral prosthesis, which is in the early stages of its development, under physiological flow conditions. In vivo magnetic resonance images (MRIs) of the left ventricle are analysed to determine the relative motion of the mitral annulus and the papillary muscle regions of the ventricle. The dynamic boundary conditions are incorporated into IB simulations to test the valve in a more realistic dynamic geometric environment. The IB model has successfully identified the effect of the dynamic boundary conditions on the mechanical behaviour of the valve and revealed the strengths and weaknesses of the current mitral design. The mechanical performance of the prosthesis is compared with recent studies of native porcine valves; differences in mechanical behaviour are observed. Potential improvements for the design of the prosthesis are proposed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.