Abstract
Ruggedness Test (RT) experiments were performed to assess the significance of the various main factors which affect pH measurements in low ionic strength aqueous solutions, as well as to establish the presence of interactions between the main factors. Stirring has an adverse effect on the measurement of pH, since it not only increases the random noise but also biases the measured value. Temperature control to the nearest 0.5 °C is sufficient for maintaining measurements accurate to 0.01 pH. Addition of NaNO3 or KCl can not be tolerated in accurate pH measurements. Three small two-factor interactions were also revealed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Research of the National Bureau of Standards
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.