Abstract

A neodymium: YAG laser operating in the thermal mode was used to irradiate isolated intraocular lenses (IOLs) and to perform transscleral cyclophotocoagulation on pseudophakic autopsy eyes to investigate the potential damage to IOL haptics such irradiation may cause. In the isolated IOLs, 70 mJ of energy deformed and partially melted both polymethylmethacrylate (PMMA) and polypropylene haptics. One of the capsular-fixated PC-IOL haptics in an autopsy eye partially melted when irradiated with the maximum energy level (8.8 J), with the aiming beam focused 1 mm posterior to the limbus and maximal posterior focus offset.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.