Abstract

Magnesium stands for a very attractive material for biodegradable stents because of its natural process and its steady disintegration into the human body by a corrosion process. The objective of the present work is to investigate the effect of the thickness on mechanical properties of the magnesium stent design. A nonlinear transient finite element simulation has been performed to analyze the influence of various thicknesses (from 50µm to 110µm with the increment of 30µm) on the behavior of a magnesium coronary stent. The model was constrained symmetrically to ensure that any virtual rigid movement does not occur during the process of coronary stent expansion. The transient load is applied in three steps in the inner surface of the stent. Four mechanical properties are studied by mathematical modeling with determination of: (1) stent deployment pressure; (2) the intrinsic elastic recoil of the material used; (3) the stent longitudinal recoil; (4) and the stress maps. The results indicate the potential application of magnesium stent and the effect of the thickness on the behavior of magnesium stent design and material.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.