Abstract
The effect of the erroneous functional forms of the slit functions on the best-fit temperature in CARS (coherent anti-Stokes Raman spectroscopy) thermometry has been investigated. When the shape of the slit function is described as a Lorentzian or a Gaussian function and its width adjusted with the room-temperature CARS spectrum, the maximum systematic errors due to the slit functions reach 1.5–2.5% in the temperature range from 300 to 2100 K. A fast-fitting algorithm exponentially interpolating the CARS spectrum at an intermediate temperature with two precalculated spectra has been proposed, and its validity has been assessed by calculating the error due to the fitting algorithm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.