Abstract
The analytical characteristics of a 27.12 MHz and a 40.68 MHz inductively coupled plasma time-of-flight mass spectrometer were evaluated under both hot and cool plasma conditions. The study was both direct and unambiguous because the same power supply, impedance-matching unit, load coil, and ICP torch were used at both operating frequencies. Although there are significant differences in the figures of merit between hot and cool plasma conditions, the performance at 27 MHz and 40 MHz was comparable. Under hot plasma conditions, measured ion energies and the plasma offset potential were nearly independent of the generator frequency. Higher ion energies and a greater plasma offset potential were apparent at 27 MHz under cold plasma conditions than at 40 MHz; however, this result may be an artifact of the difference in the optimal nebulizer flow rate rather than of frequency. Experimentally determined gas-kinetic and ionization temperatures did not demonstrate a significant frequency dependence.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.