Abstract

We present a comprehensive analysis of factors influencing the morphology of the Doppler spectrum obtained from a laser-feedback interferometer. We explore the effect of optical system parameters on three spectral characteristics: central Doppler frequency, broadening, and signal-to-noise ratio. We perform four sets of experiments and replicate the results using a Monte Carlo simulation calibrated to the backscattering profile of the target. We classify the optical system parameters as having a strong or weak influence on the Doppler spectrum. The calibrated Monte Carlo approach accurately reproduces experimental results, and allows one to investigate the detailed contribution of system parameters to the Doppler spectrum, which are difficult to isolate in experiment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call