Abstract
This paper includes simulating an electron gun design to investigate the effect of the distance between the cathode and the Wehnelt cylinder bore on the quality of the ion beam emittance and to obtain the best distance of that geometrical factor. A study of a group of designs was conducted to find this distance. The SIMION8.0 program has been used to calculate the trajectories of the electrons with initial kinetic energy of (0.1 eV) and electron current of (1×10-12 A). The investigation also includes comparisons between the equipotential line trajectories for each design as well as the field distribution and the electron trajectories. The best emittance distance was concluded to be (3 mm).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.