Abstract

The retention mechanism of acids and bases in reversed-phase liquid chromatography (RPLC) has been experimentally studied by examining the temperature dependence of retention, with emphasis on the role of the buffer ionization equilibria in the retention and selectivity. Retention factors of several ionizable compounds in a typical octadecylsilica column and using buffers dissolved in 50% (w/w) methanol as eluents at three temperatures in the range of 25–50 °C were measured. Two pairs of buffer solutions were prepared by a close adjusting of their pH at 25 °C; differences in their ionization enthalpies determined a different degree of ionization when temperature was raised and, as a consequence, a different shift in the eluent pH. Predictive equations of retention that take into account the temperature effect on both the transfer and the ionization processes are proposed. This study demonstrates the significant role that the selected buffer would have in retention and selectivity in RPLC at temperatures higher than 25 °C, particularly for co-eluted solutes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.