Abstract

Measurements using the National Synchrotron Light Source provided a detailed comparisons of the microscale detector response and infrared microscopy images for CdZnTe Frisch-ring x-ray and gamma detectors. Analysis of the data showed conclusively that local deteriorations of the electron charge collection and x-ray device response fully correlate with the presence of Te precipitates as seen in the IR images. Effects of the surface processing conditions on the detector performance were also clearly observed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.