Abstract

Regardless of the type of performed restoration, in most cases, a screw connection is employed between the abutment and implant. For this reason, implant screw loosening has remained a problem in restorative practices. The purpose of this study was to compare the surface of coated/plated screws with titanium and gold alloy screws and to evaluate the physical properties of coated/plated material after scratch tests via FE-SEM (field emission scanning electron microscopy) investigation. GoldTite, titanium screws provided by 3i (Implant Innovation, USA) and TorqTite, titanium screws by Steri-Oss (Nobel Biocare, USA) and gold screws and titanium screws by AVANA (Osstem Implant, Korea) were selected for this study. The surface, crest, and root of the abutment screws were observed by FE-SEM. A micro-diamond needle was also prepared for the scratch test. Each abutment screw was fixed, and a scratch on the surface of the head region was made at constant load and thereafter the fine trace was observed with FE-SEM. The surface of GoldTite was smoother than that of other screws and it also had abundant ductility and malleability compared with titanium and gold screws. The scratch tests also revealed that teflon particles were exfoliated easily in the screw coated with teflon. The titanium screw had rough surface and low ductility. The clinical use of gold-plated screws is recommended as a means of preventing screw loosening.