Abstract
Objective: The aim of this study is to explore the effect of stimulus frequency on air-conducted cervical and ocular vestibular evoked myogenic potential (cVEMP and oVEMP) in healthy subjects.Methods: The study included 45 healthy subjects who underwent the VEMP tests. Different stimulus frequencies (250-1500 Hz) were used for air-conducted cVEMP and oVEMP.Results: In cVEMP, P1 and N1 latencies were significantly affected by different frequencies (P < .01). The amplitude at 500 Hz was significantly larger than those at other frequencies (P < .01).There was no significant main effect of frequency on asymmetry ratio (AR) (P > .05). In oVEMP, there was a tendency for the N1 and P1 latencies to decrease from 250 Hz to 1500 Hz (P < .01). The amplitudes at 500 Hz and 1000 Hz were significantly larger than the amplitudes at 250 Hz and 1500 Hz (P < .01).There was no significant main effect of frequency on AR (P > .05).Conclusion: The optimal stimulus frequency of the cVEMP is 500 Hz and for the oVEMP is 500Hz or 1000Hz. Due to the absence of impact of stimulus frequency, AR is the best parameter of VEMP for clinical use.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.