Abstract

The risk assessment of di(2-ethylhexyl)phthalate (DEHP) migrating from polyvinyl chloride (PVC) medical devices is an important issue. Many studies have been conducted to determine the level of DEHP migration. A recent report has indicated that DEHP in blood bags is hydrolyzed by esterase into mono(2-ethylhexyl)phthalate (MEHP). However, MEHP is thought to be even more toxic than the parent compound. Therefore, a method for the simultaneous determination of DEHP and MEHP was developed. The limits of quantification (LOQs) of DEHP and MEHP were 2.5 and 0.75 ng/ml, respectively. In this study, the effect of sterilization process on the levels of DEHP and MEHP migration was investigated. The level of migration of DEHP from gamma(γ)-ray sterilized PVC sheet was low compared with that of the unsterilized control. By contrast, the level of MEHP migration from the γ-ray sterilized PVC sheet was high compared with that of the unsterilized control. In addition, a high content of MEHP was found in the γ-ray sterilized PVC sheet.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.