Abstract

AbstractThe inhomogeneity of a local magnetic field causes an image artifact of geometric distortion and intensity abnormality because of the slice offset and readout shift in slice‐selective MR imaging. It has been found that this artifact can be corrected by the projection of the slice offset onto the readout axis at a certain oblique slice angle. The slice angle for the artifact correction is determined by the amplitude of slice selection and readout gradients, and is independent of the magnetic field inhomogeneity and the main magnetic field direction. In addition, the existing view‐angle tilting technique is found to be valid only for the slice orientation orthogonal to the object axis. The slice angle effect on the inhomogeneity artifact was confirmed experimentally through phantom and volunteer's head imaging for both regular and view‐angle tilted spin echo sequences at 3 T. © 2009 Wiley Periodicals, Inc.Concepts Magn Reson Part A 34A: 238–248, 2009.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.