Abstract

Recent advances in optical technology have emerged for measuring blood flow in the extremities using speckleplethysmography (SPG), which may address needs in vascular medicine and other fields. SPG has demonstrated a highly linear response with flow rate, but the susceptibility to differences in skin tone is unclear. Two validation studies using skin-simulating phantoms and a simple clinical protocol were conducted to determine the impact of absorbing skin layers on SPG measurements. Benchtop results demonstrated that the coefficient of determination between known flow rate and SPG was highly linear (R2 = 0.990) and was unaffected by the addition of skin-phantom layers with variable absorption (R2 = 0.996-0.999). Additionally, no significant trend was found between the fit residuals of SPG and flow rate with increasing skin-phantom absorption (R2=0.025, p = 0.29). In clinical testing, no significant difference was found using both a 4-way ANOVA between demographic classifications (F = 0.89, p = 0.45), and a 2-way ANOVA test between lower- and higher-melanin subclassifications (F = 0.4, p = 0.52).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.