Abstract
The influence of electrode size on sensory nerve action potential (SNAP) amplitude of the lateral antebrachial cutaneous nerve (LACN) and sural nerve (SN) was studied in 63 healthy volunteers. The SNAP amplitudes were measured using surface recording electrodes of three different sizes, positioned across the nerve. Mean amplitudes using a 5-mm electrode were 9.0% (SN) and 15.3% (LACN) higher than with a 20-mm electrode and 19.4% (SN) and 25.8% (LACN) higher than using a 40-mm electrode. To study the influence of pressure on surface recording electrodes, studies were performed on the LACN in 31 healthy volunteers. Light pressure of the recording electrodes on the skin gave lower amplitudes (15.3%) than did greater pressure or pressure applied between active and reference electrodes. These studies demonstrate that standardized surface recording electrode size and pressure are imperative for obtaining valid and reliable results in experimental studies or in clinical follow-up of patients undergoing nerve conduction studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.