Abstract

Measurements of magnetic properties of toroidal samples of magnetically soft materials are defined by the group of international standards IEC 60404. These standards do not specify the actual value of the shunt resistor, but for instance just the fact that it should be non- inductive. This study provides more information on the topic of selection of shunt resistance, which is crucial for many studies of magnetic measurements. It was found that theresistance value of ashunt resistor has no direct effect on themeasurement resultsof magnetic properties. The quality of calibration of the value does have a direct effect, but not the value itself and it could have any value from 0.1 Ω to 10.0 Ω (as investigated). Care must be taken so that additional device-specific effects are not induced in the measurement equipment (e.g. reduced accuracy at the lowest or highest input ranges). The value of the shunt resistor should be such that the voltage drop across it would coincide with the most accurate input range of the voltmeter or data acquisition device.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.