Abstract
Circular dichroism (CD) spectroscopy is a well-established biophysical technique used to investigate the structure of molecules. The analysis of a protein CD spectrum depends on the quality of the original CD data, which can be affected by the sample purity, background absorption of the additives/solvent/buffer, the choice of the parameters used for data collection, etc. In this paper, the CD spectrum of myoglobin was used as a model to exploit how variations on each data collection parameter could affect the final protein CD spectrum and, the subsequent effect of them on the quantitative analysis of protein secondary structure. Bioinformatics analysis carried out with SESCA package and PDBMD2CD server predicted a theoretical myoglobin CD spectrum, and a Monte Carlo-like model was implemented to estimate the uncertainty in secondary structure predictions performed with CDSSTR, Selcon 3 and ContinLL algorithms. An inappropriate choice of data collection parameters can lead to a misinterpretation of the CD data in terms of the protein structural content.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.