Nowadays the magnetic field mapping is widely used in the design and analysis of the NMR/MRI magnet system, and the accuracy of mapping result has become more and more important. There are several factors affecting the accuracy of the mapping such as the mapping method, the precision of the sensor, the position of the measurement points, the calculation accuracy, and so on. In this paper the error due to the misalignment of the measurement points was discussed. The magnetic field in the central volume was mapped using an indirect method in an MRI magnet system and the magnetic field was fitted to a polynomial. Considering the misalignment between the original measurement points and the practical measurement points, there must be some errors in the mapping calculation and we called it positioning error. Several comparisons of the positioning error have been presented through the theoretical estimates and the exact magnetic field values. Finally, the allowable positioning errors were suggested to guarantee the accuracy of the magnetic field mapping within a certain degree for an example case.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.