Abstract

This study aimed to evaluate how various rotational modes influence the torque/force production and shaping ability of ProTaper Universal (PTU; non-heat-treated) and ProTaper Gold (PTG; heat-treated) nickel–titanium instruments. J-shaped resin canals were instrumented with PTU or PTG using an automated instrumentation device operated with reciprocating rotation [150° clockwise and 30° counterclockwise (R150/30) or 240° clockwise and 120° counterclockwise (R240/120)], optimum torque reverse motion (OTR), or continuous rotation (CR) (n = 10 each). Maximum force and torque were recorded, and canal centering ratios were calculated. Statistical analysis was performed with two-way ANOVA and a Bonferroni test (α = 0.05). The results were considered with reference to previous studies on the microstructure of the instruments. The upward force generated by R240/120 and OTR was smaller than that generated by R150/30 and CR in PTG (p < 0.05). The clockwise torque produced by OTR was lower than that produced by R150/30 in PTU and R240/120 and CR in PTG (p < 0.05). R240/120 and OTR induced less canal deviation compared to CR in PTU at 0 mm from the apex (p < 0.05). In conclusion, R240/120 and OTR reduced the screw-in force in PTG and improved the canal centering ability in PTU, which may be associated with the heat treatment-induced microstructural difference of the two instruments.