Abstract

The effect of reuse on dialyzer efficacy was examined by measuring blood compartment volume and dialyzer mass transfer coefficient (maximum dialyzer clearance) as a function of dialyzer use number. The 102 polysulfone dialyzers tested (F60 and HF80, Fresenius) were reprocessed on Renatron machines using peroxyacetic acid as the dual cleansing and sterilizing agent. Each dialyzer was used an average of 14.4 +/- 5.7SD times and was tested once (twice for 13/102 dialyzers) during a routine dialysis session at an arbitrary use number (7.6 +/- 5.3; range 1 to 24). The parameters tested were found to decrease only marginally with reuse, corresponding to a blood compartment volume loss of approximately 1% (R = 0.04) over a 5-week/15-use period and a decrease in dialyzer mass transfer coefficient of approximately 3% (R = 0.07 and 0.06) over the same period for urea and creatinine, respectively. It was concluded that the loss in dialyzer efficacy is negligible over the average use period of almost 5 weeks per dialyzer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.