Abstract
Sufficient dose reduction may not be achieved if radioprotective curtains are folded. This study aimed to evaluate the scattered dose rate distribution and physician eye lens dose at different curtain lengths. Using an over-couch fluoroscopy system, dH*(10)/dt was measured using a survey meter 150 cm from the floor at 29 positions in the examination room when the curtain lengths were 0% (no curtain), 50%, 75%, and 100%. The absorbed dose rates in the air at the positions of endoscopist and assistant were calculated using a Monte Carlo simulation by varying the curtain length from 0 to 100%. The air kerma was measured by 10 min fluoroscopy using optically stimulated luminescence dosimeters at the eye surfaces of the endoscopist phantom and the outside and inside of the radioprotective goggles. At curtain lengths of 50%, 75%, and 100%, the ratios of dH*(10)/dt relative to 0% ranged from 80.8 to 104.1%, 10.5 to 61.0%, and 11.8 to 24.8%, respectively. In the simulation, the absorbed dose rates at the endoscopist’s and assistant’s positions changed rapidly between 55 and 75% and 65% and 80% of the curtain length, respectively. At the 0%, 50%, 75%, and 100% curtain lengths, the air kerma at the left eye surface of the endoscopist phantom was 237 ± 29, 271 ± 30, 37.7 ± 7.5, and 33.5 ± 6.1 μGy, respectively. Therefore, a curtain length of 75% or greater is required to achieve a sufficient eye lens dose reduction effect at the position of the endoscopist.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.