Abstract

The last ten years have seen an increasing use of Coriolis flowmeters in the measurement of small flows of liquids. The performance of Coriolis meters under pulsating flow conditions was not well known, however. This paper presents some investigations on such flowmeters in the presence of monofrequent and polyfrequent pulsating flows. The linearity of the flow characteristics is checked mathematically and the mechanical resonances of the measuring tubes are determined. Two U-tube type Coriolis meters are tested and the measuring error resulting from the pulsation of the flow is examined. The investigations are carried out with monofrequent pulsation by a special test rig flow and with geared and piston type pumps (polyfrequent excitation).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.