Abstract

Primary coil dimensions can influence the accuracy of resistivity measurements made by the eddy current decay (ECD) method. Experimental results revealed that when the length of a 22.5 mm i.d. primary coil is changed from 60 to 20 mm, the corresponding measured values of resistivity increase by 4–5%. The effect occurs when there is an insubstantial background field at the sample ends. ECD measurement accuracy is better than 1% when the primary coil length to inside diameter ratio is larger than 2.2 and the sample diameter is 44% or less of the primary coil i.d. The investigation included measurements at room temperature, 77 K and 4.2 K on samples having residual resistivity ratios of 153, 340, 1010 and 7790.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.