Abstract

The conventional ferrous xylenol orange (XO) gel (FXG) dosimeter is being wildly investigated for radiotherapy dose measurements. Upon irradiation, its color turns red due to oxidation of Fe2+ into Fe3+, which forms a complex with xylenol orange. The effect of perchloric acid (PCA) on the dosimetric properties of the gel in the dose range of 1–15Gy was investigated using visual spectrophotometry. FXG-PCA responds to radiation dose linearly and exhibits higher radiation sensitivity than the conventional gel dosimeter. PCA in a concentration of 20mM enhances the radiation sensitivity ~44%. Stability of the absorbances of both the gels during storage under various conditions was investigated, and the uncertainty of dose measurements was estimated.