Abstract
The high performance liquid chromatography (HPLC) separation methodology employed in the study of polyalkene additive compounds by atmospheric pressure ionization mass spectrometry (API-MS) was undertaken. Both atmospheric pressure chemical ionization (APCI) and atmospheric pressure photoionization (APPI) were examined. APPI (including dopant-assisted APPI) was found to be an inferior ionization technique to APCI in all cases. APCI ion responses were found to be highly dependent upon the organic solvent type used in the HPLC separations. Namely, employing a water/methanol gradient in place of a water/acetonitrile or a water/acetone gradient yielded improvements in analyte ion intensities between 2.3- and 52-fold for the liquid chromatography–mass spectrometry (LC–MS) experiments. Analyte and mobile phase solvent ionization energies were found to be only partially responsible, whereas mobile phase cluster formation and hydration was also implicated. Mobile phase component modification is demonstrated to be an important consideration when developing new, or modifying existing HPLC separations for use in LC–MS experiments in order to enhance analyte sensitivity for a wide variety of common polyalkene additives.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.