Abstract

X-ray phase-contrast and dark-field imaging are two new modalities that have great potential for applications in different fields like medical diagnostics or materials science. The use of grating interferometers allows the detection of both differential phase shift and dark-field signal together with the absorption signal in a single acquisition. We present wave-optical simulations to quantitatively analyze the response of a grating-based X-ray phase-contrast and dark-field imaging setup to variations of the sample relative to the system. Specifically, we investigated changes in the size and the position of the object. Furthermore, we examined the influence of different detector pixel sizes while sample and interferometer remained unchanged. The results of this study contribute to a better understanding of the signal formation and represent a step towards the full characterization of the response of grating interferometry setups to specific sample geometries.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.