Abstract

Ultrasonic pulse Doppler method has been widely used in many engineering field referred to as UVP (ultrasonic velocity profiler). Analysis algorithm in the UVP, number of pulse repetition (Npulse), noise and reflector conditions, etc. may affect on the measurement accuracy. The Npulse is related to the temporal resolution, and must be set as low as possible in order to improve the temporal resolution. However, it is known that accuracy of instantaneous velocity becomes worse with decreasing of the Npulse. In this study, effect of analysis algorithm in the UVP on velocity data was investigated with changing of the Npulse and SNR by simulation and experiments. As a result, it is shown that there is an appropriate Npulse in each algorithm depending on the SNR. An algorithm which decides the Doppler frequency from maximum value in the power spectrum, FFT-max, was relatively hard to be affected by noise. Difference of velocity standard deviation was small between lower and higher noise conditions when FFT-max with Npulse over the appropriate value was used for the calculation. Hence, FFT-max was the best analysis algorithm for measuring flow field compared with the autocorrelation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.