Abstract

Concentrations of halogenated anesthetics produced by contemporary vaporizers vary from vaporizer dial settings when carrier gas is not 100% oxygen. This effect is most marked when carrier gas changes from 100% O2 to 100% nitrous oxide (N2O). Vapor concentrations of halothane and enflurane from vaporizers with 6 L/min carrier gas flows were reduced to 3% from 4% until absorption of N20 by the liquid anesthetic ceased. Presaturation of liquid anesthetics with N2O eliminated the transient decrease of vapor output. Steady-state outputs of halothane and enflurane in 100% N2O were 10% below dial settings. The significance of these changes in administration of closed-circuit anesthesia with an out-of-circuit vaporizer is discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.