Abstract

BackgroundThis study aimed to evaluate the intrarenal pelvic pressure in endoscopic combined intrarenal surgery using an artificial kidney model.MethodsAn artificial kidney model was created using the Urovac evacuator™. Four sizes of nephrostomy sheaths (MIP-L: 25/26 Fr, MIP-M: 16.5/17.5 Fr, MIP-S: 11/12 Fr, MIP-XS: 8.5/9.5 Fr) and two sizes of ureteral access sheaths (12/14 Fr and 10/12 Fr) were installed into the model. For each combination of nephrostomy and ureteral access sheath, renal pelvic pressure was measured with and without insertion of the retrograde flexible ureteroscope. Irrigation from the nephroscope was adjusted to 40–160 mmHg using an automatic irrigation device, and the irrigation of the ureteroscope was by spontaneous dripping at 80 cmH2O. Conditions were measured six times, and the renal pelvic pressure was compared in different conditions.ResultsWithout ureteroscope insertion through the ureteral access sheath, the renal pelvic pressure never exceeded 30 mmHg. Meanwhile, when the ureteroscope was inserted, the renal pelvic pressure increased as the nephrostomy sheath and ureteral access sheath became narrower and as the irrigation pressure increased. Intrarenal pelvic pressure exceeded 30 mmHg when the irrigation pressure was increased in 12/14 Fr ureteral access sheath when MIP-XS was used, and in 10/12 Fr ureteral access sheath when MIP-XS and MIP-S were used.ConclusionsThe use of a thin nephrostomy sheath in endoscopic combined intrarenal surgery can lead to increased intrarenal pelvic pressure. Although our results are from an artificial kidney model, special care is suggested to be required when using a retrograde flexible ureteroscope simultaneously in treatment of patients.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call