Abstract

Purpose: To compare image quality and radiation exposure between super- and ultra-high-resolution helical and super-high-resolution volumetric CT of the temporal bone. Methods: Six cadaveric temporal bone specimens were used to evaluate key temporal bone structures using the following CT reconstruction and acquisition modes: helical and single-volume acquisition modes in super-high resolution (0.25-mm slice thickness, 10242 matrix), and helical mode in ultra-high resolution (0.25-mm slice thickness, 20482 matrix). Two observers performed 5 previously described preoperative measurements, measured noise and signal-to-noise ratios for air, and noise for bone, and rated the visualization of 5 anatomical structures on a 4-point scale, for each reconstruction mode. Radiation dose exposure was recorded for each examination. Results: There was no significant difference between any of the quantitative or qualitative measurements in any of the reconstruction and acquisition modes. There was a slight increase in noise and a decrease in signal-to-noise ratio in the air using the single-volume mode (115 ± 13.1 HU and 8.37 ± 0.91, respectively) compared to the helicoidal super-high-resolution (92.4 ± 11.8 HU and 10.8 ± 1.26, respectively) and helicoidal ultra-high-resolution (91.1 ± 10.7 HU and 10.9 ± 1.39, respectively) modes ( P < .002). The volumic CT dose index was 50.9 mGy with helical acquisition and 29.8 mGy with single-volume acquisition mode ( P < .0001). Conclusion: The single-volume super-high-resolution acquisition mode allows a reduction in radiation dose exposure without compromising image quality compared to helical scanning, but with a slightly lower signal-to-noise ratio in air with the single-volume mode, while there was no difference in image quality between the helical super- and ultra-high-resolution modes.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call