Abstract

Three groups of 15 DMEK grafts (five per group) were pre-loaded in the DMEK RAPID Mini and preserved in Optisol-GS for 72 h at different temperatures: group A at >8 °C, group B between 2-8 °C and group C at <2 °C. After stripping and preservation, the viability of the endothelium, cell loss and morphology were assessed through light microscopy following trypan blue and alizarin red staining. Overall mortality was 4.07%, 3.97% and 7.66%, in groups A, B and C, respectively, with percentages of uncovered areas of 0.31%, 1.36% and 0.20% (all p > 0.05). Endothelial cell density variation was 5.51%, 3.06% and 2.82% in groups A, B and C, respectively (p = 0.19). Total Endothelial Cell Loss (ECL) was 4.37%, 5.32% and 7.84% in groups A, B and C, respectively (p = 0.39). Endothelial cell morphology was comparable in all three groups. In the DMEK RAPID Mini, low temperatures (<2 °C) may affect the quality of pre-loaded grafts, inducing a higher ECL after 72 h of preservation, although no significant differences among groups could be proved. Our data would suggest maintaining grafts loaded in the DMEK RAPID Mini at temperatures between 2-8 °C for appropriate preservation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.