Abstract

Nd-YAG laser pulses of 1064, 532, 266nm in wavelength and of 8ns and 100ps in pulse width as well as KrF excimer laser pulses (248nm) of 30ns pulse width were employed to deposit thin films of the simulated spent nuclear fuel pellets. The deposited thin films were dissolved in nitric acid and analyzed for residual heavy element (U) to burn-up monitor (Nd) ratio using isotopic dilution mass spectrometry (employing the TIMS technique) to determine the burn-up in these simulated fuel pellets. While the use of ns-laser pulses of 1064nm wavelength resulted in high elemental fractionation with enhanced Nd concentration compared to the target composition, the 266nm ps-laser pulses produce films with least elemental fractionation in the film and hence is best suited for this analytical application. Due to the very small amount of material in the deposited film and the amenability of pulsed laser deposition system for use in radioactive shielded hot cells, this analytical technique is suitable for sampling highly radioactive materials.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.