Addition of a small amount of polar solvent (i.e., modifier) to CO 2 in packed column supercritical fluid chromatography (SFC) has shown major improvements in both polar analyte solubility and interaction of the polar analyte with the stationary phase. Recently, the addition of an ionic component (i.e., additive) to the primary modifier by one of us has been shown to extend even further the application of SFC to polar analytes. In this work, the effect of various ionic additives on the elution of ionic compounds, such as sodium 4-dodecylbenzene sulfonate and sodium 4-octylbenene sulfonate, has been studied. The additives were lithium acetate, ammonium acetate, tetramethylammonium acetate, tetrabutylammonium acetate, and ammonium chloride dissolved in methanol. Three stationary phases with different degrees of deactivation were considered: conventional cyanopropyl, deltabond cyanopropyl, and bare silica. The effect of additive concentration and additive functionality on analyte retention was investigated. Sodium 4-dodecylbenzene sulfonate was successfully eluted using all the additives with good peak shape under isocratic/isobaric/isothermal conditions. Different additives, however, yielded different retention times and in some cases different peak shapes.