Abstract
In some earlier hip simulator studies of large-diameter metal-on-metal (MoM) hip prostheses high, unexplained variation in wear has been observed. In the present study it was found that high variation occurs when the tribological endurance limit of the MoM device in question is being approached. When the limit is exceeded, high wear takes place uniformly, and the variation is low. Below the limit, both the wear and its variation are low. The endurance limit was exceeded by increasing the peak load from 2kN to 3kN, to simulate obesity. The acetabular cup inclination angle was high, 70°. Potential clinical wear problems can be predicted by hip simulator studies provided that relevant adverse test conditions are included.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.