Abstract

The objective of this study was to evaluate the effect of the increased instantaneous input dynamic range (IIDR) in the Nucleus Freedom cochlear implant (CI) system on recipients' ability to perceive soft speech and speech in noise. Ten adult Freedom CI recipients participated. Two maps differing in IIDR were placed on each subject's processor at initial activation. The IIDR was set to 30 dB for one map and 40 dB for the other. Subjects used both maps for at least one month prior to speech perception testing. Results revealed significantly higher scores for words (50 dB SPL), for sentences in background babble (65 dB SPL), and significantly lower sound field threshold levels with the 40 compared to the 30 dB IIDR map. Ceiling effects may have contributed to non-significant findings for sentences in quiet (50 dB SPL). The Freedom's increased IIDR allows better perception of soft speech and speech in noise.