Abstract

Wear and corrosion at the modular head-neck junction has been recognised to be a potential clinical concern, with multiple reports on adverse local tissue reactions and subsequent early failure of metal-on-metal hip replacements. Furthermore, reports on head-neck taper corrosion are also being described with conventional metal-on-polyethylene bearings. Manufacturing tolerances, surgical technique, non-axial alignment, material combination, high frictional torque and high bending moment have all been implicated in the failure process. There is limited guidance on the force of impaction with which surgeons should assemble modular hip prostheses. This study aims to investigate the effect of impaction force on the deformation and corrosion of modular tapers. Short neck tapers with high surface roughness (average Rz = 16.58 μm, Ra = 4.14μm) and long neck tapers with low surface roughness (average Rz = 3.82 μm, Ra = 0.81μm), were assembled with CoCrMo alloy heads (smooth finish) under controlled conditions with 2, 4 or 8 kN of impaction force. Material combinations tested included CoCrMo-head/CoCrMo-neck and CoCrMo-head/Ti-6Al-4V-neck. Assessment of surface deformation before and after impaction was made using surface profilometry. Measurement of fretting current during sinusoidal cyclic loading evaluated mechanically assisted corrosion for each assembly load during short-term cyclic loading (1000-cycles) and long-term cyclic loading (5 million-cycles). Deformation on head and neck tapers increased with assembly load. Fretting currents during short term simulation testing showed significantly lower currents (p < 0.05), in 8 kN assemblies when compared to 2 and 4 kN, especially for the short-rough tapers. Long-term simulator testing demonstrated a progressive reduction in fretting corrosion for samples impacted with 4 and 8 kN; however, this reduction was greater for samples impacted at 8 kN even at the start of testing. Based on our results, surgeons could minimise mechanically assisted crevice corrosion by using higher impact loads when assembling the head to the stem in total hip arthroplasty. © 2017 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 36:405-416, 2018.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.